Contact an Account ManagerDetermination of selected polyfluorinated compounds (PFC) in water acc. to DIN 38407-42
Summary
Type: HPLC
Application #: 126530
Detection:
Conditions: Eluent A: 10 mmol/L ammonium acetate in water – methanol (75:25, v/v)
Eluent B: 10 mmol/L ammonium acetate in acetonitrile – methanol (75:25, v/v)
Gradient: 0–30 % B in 5 min, 30–55 % B in 8 min, 55–80 % B in 19 min
Flow rate: 0.3 mL/min
Temperature: 50 °C
Injection volume: 10 µL (each concentration: 10 ng/mL)
Authors: Normenausschuss Wasserwesen (NAW) im DIN
Source: DIN 38407-42:2011-03